Abstract

A small test ion chamber has been constructed using 2,2,4,4 Tetramethylpentane as the ionizing liquid in direct contact with an array of lead plates used as the high voltage and ground planes. The transient response resulting from the passage of cosmic ray muons through the ion chamber was monitored for a period of 54 weeks to measure possible degradation of the signal due to contamination of the ionizing liquid that may have come from the lead plates. Over the first 17 weeks of the experiment, the loss of pulse height was measured to be less than 11% per year at 95 confidence. During the following 37 weeks, a more sensitive measurement was consistent with zero loss, or a limit of less than 2.9% per year at 95% confidence.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.