Abstract

The construction of plasticised PVC matrix-type scopolamine ion- selective membrane electrode and its use in the potentiometric determination of scopolamine in pharmaceutical preparations are described. It is based on the use of tlie ion-associate species, fonned by scopolainine cation and 5-nitrobarbitoric counter ion. The basic electrode peifonnance characteristics are evaluated according to IUPAC I-ecoininendatioiis. It exhibited a linear response for 1x10-2–1x10-5 M of scopolamine solutions with a cationic Nernstian slope over the pH range of 4-7. Common organic and inorganic cations showed negligible interference. Direct potentiometric determination of 1x10-2–1x10-5 M aqueous tubocurarine chloride using this membrane electrode system showed an average recovery of 99.05 with a mean standard deviation of 0.12. This electrode system was successfully applied to the potentiometric determination of scopolamine in some pharmaceutical preparations.

Highlights

  • The coiistniction of plasticised PVC matrix-type scopolamine ionselective ineinbrane electrode and its use in tlie potentioinetric detenninatioii of scopolainine in pliannaceutical preparations are described

  • Aqueous tuboctirarine chloride using this ineinbrane electrode systein sliowed ail average recovery of 99.05 witli a inean standard deviation of 0.I. This electrode system was successfillly applied to tlie potentioinetric detenninatioii of scopolainine in some pliannaceutical preparations

  • A inirt~lreof e q ~ ~ avol lumes of 5x10-.' M scopolainine bu~tylbromide and 5r10-' M potassium chloride was used as internal reference solution in \vIiich AgIAgCI reference electrode is dipped

Read more

Summary

Introduction

The coiistniction of plasticised PVC matrix-type scopolamine ionselective ineinbrane electrode and its use in tlie potentioinetric detenninatioii of scopolainine in pliannaceutical preparations are described. Sensor Preparation: The scopola~nine-5-nitrobarbituratewas prepared by mixing 20 1n1 of I M aqueous scopolamine butyl bromide with 20 in[1] of 1 M ethanolic 5nitrobarbituric acid solution '47'. Electrode Calibration: Aliquots of 10 in[1] of l slo-' - 1x 10"' M scopolainine butyl broinide solution were transferred into 50 in[1] beakers and the PVC scopolamine-5nitrobarbitlrrate ineinbrane electrode in conjilnction with an Orion double-jt~~ictioAigi -AgCl reference electrode (Model 90-02) were immersed in the soliltion.

Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call