Abstract

We attempted to construct and evaluate a novel detection method to realize simultaneous detection based on a multiplex liquid protein chip technique for nine nutrition-and-health-related protein markers to meet the requirement of an accurate, simultaneous and comprehensive analysis of the proteomics of nutrition and health. The lower limits of detection, biological limits of detection and regression equations of serum ferritin (SF), soluble transferrin receptor (sTfR), c-reactive protein (CRP), retinol-binding protein4 (RBP4), apolipoprotein B (ApoB), alpha-fetoprotein (AFP), prealbumin (PA), carcino-embryonic antigen (CEA) and D-Dimmer (D-D) were determined after a series of optimal experiments. Then, the results of the methodological evaluation for this novel method indicated that the accuracies were between 70.12% and 127.07%, the within-run precisions were between 0.85% and 7.31%, the between-run precisions were between 3.53% and 19.07%, the correlation coefficients between this method and other methods were above 0.504 (p < 0.05), and the direct bilirubin (DBIL) of low concentration and the indirect bilirubin (IBIL) of high concentration could not interfere with the detected results of nine indicators. The novel multiplex detection method, which can increase accuracy and improve the ability of comprehensive analysis, can basically meet the requirement of detection and the diagnosis of the proteomics of nutrition and health.