Abstract

The development and application of a constrained Single Input Single Output (SISO) version of the popular Generalised Predictive Control (GPC) algorithm, which uses the Quadratic programming (QP) approach, is presented in this paper; Mean Arterial pressure (MAP) is used as an inferential variable to indicate the level of unconsciousness. First, the algorithm was validated using a derived re-circulatory physiological model of anaesthesia via a semi-closed circuit before the closed-loop control system was transferred to the operating theatre for validation during surgical operations. Simulation and real-time experiments showed that excellent regulation of blood pressure around set-point targets can be achieved. Such regulation was later translated to being equivalent to a good maintenance of level of anaesthesia.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.