Abstract
A constant-fraction discrimination (CFD) system has been combined with a boxcar integrator for detection in inductively coupled plasma/time-of-flight mass spectrometry. The discriminator provides gating logic for the boxcar integrator when an incoming ion signal occurs, but discriminates against electronic or background noise of lower amplitude. As a result, the combination can effectively reject noise and accumulate analyte signal, rather than relying on an averaging process to reduce noise levels. The signal-to-noise ratio is therefore enhanced in this operation compared with the conventional boxcar method. The dynamic range of the detection system is at least five orders of magnitude.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.