Abstract

Constant-current enhanced potentiometric stripping analysis (CCEPSA) is similar to potentiometric stripping analysis (PSA) except that during the stripping step in CCEPSA a cathodic current is imposed on the electrode system. The freshly oxidized metal ions in the vicinity of the electrode surface are partially redeposited so that the stripping time is extended. The stripping time and potential-time equations for CCEPSA were derived and verified using the stripping of cadmium(II). Under the most favorable conditions, the sensitivity of cadmium(II) determination was enhanced by a factor of 15.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.