Abstract
This work reports that the intensity ratio of spectral lines having similar self-absorption characteristics during laser induced breakdown spectroscopy (LIBS) analysis can become nearly constant over a wide range of irradiation conditions if the intensities are integrated over a sufficiently long time. It is shown that the plasma temperature and intensity ratio of these spectral lines have temporal similarity. The spectral lines with similar self-absorption properties may be selected to improve the accuracy and consistency of LIBS analysis results under an environment with fluctuating measurement conditions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.