Abstract

Background and purpose: To gain insight into the current practice of quality control (QC) of medical electron accelerators and to reduce possible variations in test frequencies and test procedures. Materials and methods: An extensive questionnaire on QC procedures of medical electron accelerators was distributed and completed by all (21) radiotherapy institutions in The Netherlands. The questions were related to safety systems, mechanical parameters, beam profiles, beam energy, absolute dosimetry, wedge filters, the dose monitor system and radiation leakage. Results: The data of the questionnaire were compared with recommendations given in national and international reports on QC of electron accelerators. Large variations in time spent on QC exist, especially for accelerators having dual energy photon beams and several electron beam energies. This diversity is mainly due to differences in philosophy with regard to QC and the differences in resources and machine time available. Furthermore, large variations in test frequencies and test methodologies were observed. The staffing level involved in the QC measurements was evaluated and compared with recent recommendations provided by EFOMP-ESTRO. Conclusions: From these recommendations and the results of the questionnaire, a set of minimum guidelines for a QC programme could be formulated and implemented in all radiotherapy institutions in The Netherlands.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.