Abstract
One of the major considerations in siting choice for an MRI system within a hospital is its interactions with its environment. This interaction places restrictions on the proximity of equipment sensitive to magnetic field and limits areas of general public access. In addition special account must be taken of the possible impact of environmental iron on magnet homogeneity. To date, the approach adopted by the MRI system scanner manufacturers to these problems has been to employ either YOKE or MIRROR iron (PASSIVE) shielding frequently requiring significant structural modifications with associated costs. An alternative approach to the shielding problem has been investigated at OXFORD using a geometry utilising superconducting counter running coils alone and prototypes with Central fields of 0.5T, 1.0T and 1.5T have been tested.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.