Abstract
The principles of the ISO "Guide to the expression of uncertainty in measurement" are applied to the analysis of solid materials by graphite furnace atomic absorption spectrometry. Assessment of trueness is achieved by the analysis of a certified reference material (CRM) under fixed instrumental conditions. For the evaluation of the method's uncertainty (as part of the validation) a model equation is derived, considering all significant contributions, including the uncertainty in the CRM value and the uncertainty in the CRM analysis. A concrete example is presented and discussed (soya flour as an unknown sample and CRM BCR rye grass). For routine analysis a pooled estimate from the validation experiment can be used, leading to an acceptable small uncertainty range even in the case of "duplicate determination". The test for trueness (acceptance) of the result of the CRM analysis is applied and the correction for detected bias is discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.