Abstract

The effects caused by an excess quantity of ionized gas molecules within the low vacuum, variable pressure and environmental scanning electron microscope (ESEM) are described with reference to mechanisms by which they can influence imaging conditions. These effects can include specimen charging, recombination and development of space charge. They are demonstrated for three different classes of sample: (1) an electrically grounded conductor, (2) an electrically floating conductor, and (3) an electrical insulator. A new device is presented that will aid excess charge removal within the ESEM and help correct for some of these effects, thereby dramatically improving imaging over a wide range of operating conditions and samples. The mechanism of image enhancement is demonstrated with reference to the three classes of sample described above.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.