Abstract

Purpose. To compare the reliability of five different intraoral scanners (IOSs) in the capture of implant scanbodies (SBs) and to verify the dimensional congruence between the meshes (MEs) of the SBs and the corresponding library file (LF). Methods. A gypsum cast of a fully edentulous maxilla with six implant analogues and SBs screwed on was scanned with five different IOSs (PRIMESCAN®, CS 3700®, MEDIT i-500®, ITERO ELEMENTS 5D®, and Emerald S®). Ten scans were taken for each IOS. The resulting MEs were imported to reverse engineering software for 3D analysis, consisting of the superimposition of the SB LF onto each SB ME. Then, a quantitative and qualitative evaluation of the deviations between MEs and LF was performed. A careful statistical analysis was performed. Results. PRIMESCAN® showed the highest congruence between SB MEs and LF, with the lowest mean absolute deviation (25.5 ± 5.0 μm), immediately followed by CS 3700® (27.0 ± 4.3 μm); the difference between them was not significant (p = 0.1235). PRIMESCAN® showed a significantly higher congruence than MEDIT i-500® (29.8 ± 4.8 μm, p < 0.0001), ITERO ELEMENTS 5D® (34.2 ± 9.3 μm, p < 0.0001), and Emerald S® (38.3 ± 7.8 μm, p < 0.0001). CS 3700® had a significantly higher congruence than MEDIT i-500® (p = 0.0004), ITERO ELEMENTS 5D® (p < 0.0001), and Emerald S® (p < 0.0001). Significant differences were also found between MEDIT i-500® and ITERO ELEMENTS 5D® (p < 0.0001), MEDIT i-500® and Emerald S® (p < 0.0001), and ITERO ELEMENTS 5D® and Emerald S® (p < 0.0001). Significant differences were found among different SBs when scanned with the same IOS. The deviations of the IOSs showed different directions and patterns. With PRIMESCAN®, ITERO ELEMENTS 5D®, and Emerald S®, the MEs were included inside the LF; with CS 3700®, the LF was included in the MEs. MEDIT i-500® showed interpolation between the MEs and LF, with no clear direction for the deviation. Conclusions. Statistically different levels of congruence were found between the SB MEs and the corresponding LF when using different IOSs. Significant differences were also found between different SBs when scanned with the same IOS. Finally, the qualitative evaluation revealed different directions and patterns for the five IOSs.

Highlights

  • Digital technologies are revolutionising the world of dentistry [1]

  • Our present in vitro study aimed to compare the reliability of five different intraoral scanners (IOSs) (PRIMESCAN®, CS 3700®, MEDIT i-500®, ITERO ELEMENTS 5D®, and Emerald S®) in the capture of implant SBs and to verify the dimensional congruence between the MEs of the SBs and the corresponding library file (LF)

  • At the end of the quantitative evaluation, statistically different levels of congruence were found between the SB MEs captured with the different IOSs and the corresponding LF

Read more

Summary

Introduction

The digital prosthetic workflow is divided into four phases: the acquisition of data with a scanner, which allows obtaining a mesh (ME), i.e., a surface reconstruction of the scanned model; the processing of the ME within CAD software, for designing the prosthetic restorations; the fabrication of the restorations by milling or 3D printing; and the clinical application [1,5,9]. All these steps determine the quality of the clinical result [9,10]. The technician can model all the restorations, which will be fabricated and applied clinically

Objectives
Methods
Findings
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.