Abstract

Laser Induced Fluorescence (LIF) provides a non-invasive, quantitative measurement of fluorescent tracers in complex mixing flows. The common availability of broad linewidth lasers and their use with iodine vapor tracers raises a significant concern about confounded tracer concentration measurements, due to the strong dependence of iodine fluorescence on both tracer concentration and flow temperature. We model the sensitivity of iodine fluorescent response to temperature fluctuations, especially for flows with significant temperature contrasts to ambient conditions, and establish a temperature uniformity requirement that bounds the relative error in tracer concentration measurements. We show that in a conventional laboratory setup, in order to bound the measured concentration error to about 5%, the relative temperature fluctuations in iodine LIF must be approximately an order of magnitude smaller than the desired measurement resolution of relative concentration fluctuations. Iodine LIF should thus be employed with great caution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.