Abstract

In this paper, we propose a confocal microscopy based on dual blur depth measurement (DBCM). The first blur is defocus blur, and the second blur is artificial convolutional blur. First, the DBCM blurs the defocus image using a known Gaussian kernel and calculates the edge gradient ratio between it and the re-blurred image. Then, the axial measurement of edge positions is based on a calibration measurement curve. Finally, depth information is inferred from the edges using the original image. Experiments show that the DBCM can achieve depth measurement in a single image. In a 10×/0.25 objective, the error measured for a step sample of 4.7397µm is 0.23µm. The relative error rate is 4.8%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.