Abstract

A detailed model of a FeMRI gel head phantom has been developed to verify the 3D radiation dose mapping of conformal therapy treatment planning. The phantom consists of a human skull filled with a solution that is 1 mM in ferrous sulfate, 1 mM in sodium chloride, and 0.1 N in sulfuric acid, gelled with 7.5% by weight gelatin. The phantom was irradiated and the absorbed dose from multiple non‐coplanar fields was measured using the MRI dose‐dependent variation of ferric sulfate. These studies were carried out using a 1.5 T MR imager and detailed analysis was performed on a UNIX workstation. The phantom and the calibration vials were scanned simultaneously using a multiple‐scan inversion recovery sequence. The values of were evaluated on a pixel‐by‐pixel basis for entire volume through the use of custom‐built software on the workstation and were converted to dose using calibration data. Comparisons of dose distributions between those measured by FeMRI and those calculated by 3D treatment planning show good agreement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.