Abstract
We evaluated the use of the INNO-LIA HIV-1/HIV-2 Ab test (LIA HIV; Innogenetics) for the confirmation of antibodies to human immunodeficiency virus type 1 (HIV-1) and type 2 (HIV-2). The test includes three recombinant HIV-1 proteins: p24 (gag), p17 (gag), and endonuclease (p31; pol), in combination with two synthetic peptides derived from the env gene of HIV-1 and one synthetic peptide selected from the env gene of HIV-2. Analysis of 450 sera from blood donors, 220 sera from patients with non-HIV pathology, and 28 Western blot (WB) p24-only reactive sera revealed no false-positive results, and the rate of indeterminate results was substantially lower than that with WB. Testing of 334 WB-confirmed HIV antibody-positive sera (309 HIV-1; 25 HIV-2) revealed no false-negative results. In two of seven seroconversion panels tested, LIA HIV detected the presence of HIV antibodies before WB did. In the other five panels, LIA HIV and WB confirmed the presence of HIV antibodies in the same sample. The LIA HIV assay therefore appears well suited for routine confirmation of the presence of HIV-1 and HIV-2 antibodies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.