Abstract
Robotic control of flexible devices can enhance and simplify many medical procedures. We present a method for controlling a tendon-driven continuum manipulator by means of specifying the shape configuration. The basis for control is a linear beam configuration model that transforms beam configuration to tendon displacement by modeling internal loads of the compliant system. An essential aspect of this model is the inclusion of both the mechanical and geometrical coupling among serial articulating sections. Important capabilities of this model are the general forward kinematics and the decoupled inverse kinematics that allow for independent control of multiple sections. Tracking results are presented for a cardiac catheter with two articulating sections.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.