Pickup coils are widely used for measurements of magnetic susceptibility and hysteretic losses in magnetic and superconducting materials. A general formulation for the calibration of such pickup coils with different geometries is presented. Although the procedure described here is general and can be applied to any coil and sample, special emphasis has been placed on the calibration of saddlelike coils usually used to measure ac losses of long superconductors under ac fields perpendicular to their long dimension. Numerical simulations and experimental measurements have been carried out in order to determine the geometrical limitations of the pickup coil/sample configurations. This is especially important when simple coils of small dimensions and number of turns are used and/or samples of high demagnetization factor are measured. Superconductors of different geometry, in particular, wires and tapes under parallel and perpendicular fields, have been analyzed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.