Abstract
BackgroundKnowledge of the cochlear implant array’s precise position is important because of the correlation between electrode position and speech understanding. Several groups have provided recent image processing evidence to determine scalar translocation, angular insertion depth, and cochlear duct length (CDL); all of which are being used for patient-specific programming. Cone beam computed tomography (CBCT) is increasingly used in otology due to its superior resolution and low radiation dose. Our objectives are as followed:Validate CBCT by measuring cochlear metrics, including basal turn diameter (A-value) and lateral wall cochlear duct length at different angular intervals and comparing it against microcomputed CT (uCT).Explore the relationship between measured lateral wall cochlear duct length at different angular intervals and insertion depth among 3 different length electrodes using CBCT.MethodsThe study was performed using fixed human cadaveric temporal bones in a tertiary academic centre. Ten temporal bones were subjected to the standard facial recess approach for cochlear implantation and imaged by CBCT followed by uCT. Measurements were performed on a three-dimensional reconstructed model of the cochlea. Sequential insertion of 3 electrodes (Med-El Flex24, 28 and Soft) was then performed in 5 bones and reimaged by CBCT. Statistical analysis was performed using Pearson’s correlation.ResultsThere was good agreement between CBCT and uCT for cochlear metrics, validating the precision of CBCT against the current gold standard uCT in imaging. The A-value recorded by both modalities showed a high degree of linear correlation and did not differ by more than 0.23 mm in absolute values. For the measurement of lateral wall CDL at various points along the cochlea, there was a good correlation between both modalities at 360 deg and 720 deg (r = 0.85, p < 0.01 and r = 0.79, p < 0.01). The Flex24 electrode displayed consistent insertion depth across different bones.ConclusionsCBCT reliably performs cochlear metrics and measures electrode insertion depth. The low radiation dose, fast acquisition time, diminished metallic artifacts and portability of CBCT make it a valid option for imaging in cochlear implant surgery.
Highlights
Cochlear implantation (CI) is a significant breakthrough in the auditory rehabilitation of patients with bilateral severe to profound hearing loss [1, 2]
Ten temporal bones were subjected to the standard facial recess approach for cochlear implantation and imaged by Cone beam computed tomography (CBCT) followed by uCT
For the measurement of lateral wall cochlear duct length (CDL) at various points along the cochlea, there was a good correlation between both modalities at 360 deg and 720 deg (r = 0.85, p < 0.01 and r = 0.79, p < 0.01)
Summary
Cochlear implantation (CI) is a significant breakthrough in the auditory rehabilitation of patients with bilateral severe to profound hearing loss [1, 2]. There has been an expansion of implant candidacy to include patients with residual hearing In these patients, the use of electroacoustic stimulation mandates that low frequency regions of the cochlea are preserved anatomically and functionally to facilitate acoustic stimulation, while the high frequency regions are stimulated electrically [3]. As there is a known variation in cochlear duct length (CDL), knowledge of the anatomy of the cochlear duct and position of the CI electrode in these patients will allow implant surgeons to customize electrode choice and insertion depth to improve functional outcomes [2, 6]. 2. Explore the relationship between measured lateral wall cochlear duct length at different angular intervals and insertion depth among 3 different length electrodes using CBCT
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Otolaryngology - Head & Neck Surgery
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.