Abstract

Imaging parameters, frequencies and resulting patient organ doses in treatments of prostate cancer were assessed in Finnish radiotherapy centres. Based on a questionnaire to the clinics, Monte Carlo method was used to estimate organ doses in International Commission on Radiological Protection standard phantom for prostate, bladder, rectum and femoral head. The results show that doses from cone beam computed tomography imaging have reduced compared to earlier studies and are between 3.6 and 34.5mGy per image for the above-mentioned organs and for normal sized patients. There still is room for further optimization of the patient exposure, as many centres use the default imaging parameters, and the length of the imaged region may not be optimal for the purpose.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.