Abstract

This paper presents an analysis of two cone beam configurations (having focal lengths of 40 and 60 cm) for the acquisition of single photon emission computed tomography (SPECT) projection data. A three-dimensional filtered backprojection algorithm is used to reconstruct SPECT images of cone beam projection data obtained using Monte Carlo simulations. The mathematical analysis resulted in on-axis point source sensitivities (calculated for a distance of 15 cm from the collimator surface) for cone beam configurations that were 1.4-3 times the sensitivities of parallel-hole and fan beam geometries having similar geometric resolutions. Cone beam collimation offers the potential for improved sensitivity for SPECT devices using large-field-of-view scintillation cameras.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.