Abstract

<span lang="EN-US">This paper investigates the effect of two different heating power control systems of infant incubators on their conducted emissions. Two infant incubators which respectively employ zero-crossing control mode and phase angle control mode are observed. The research was conducted by measuring conducted electromagnetic interference (EMI) from each infant incubator's power input. Measurements are conducted both during full power condition, while the incubator's compartment temperature is far away from the temperature setpoint, and during power chopping condition, while the compartment temperature reaches the steady-state set point. Method and limit of the measurement refer to CISPR 11. It is found that conducted emission higher than the standard CISPR 11 limit occurs during power chopping on phase angle control mode. This results from the sharp rise time of voltage delivered to the heater, around 220 ns for each chopping cycle.</span>

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.