Abstract

The amperometric determination of chemical oxygen demand (COD) reported by Quan Xie’s group (Electrochem Commun 9:2281, 14), was a rapid, green and simple COD evaluation method. This work focused on testing and verifying this method by using a home-made boron-doped diamond (BDD) film as anode and optimizing the experiment conditions. The BDD thin film electrode was employed as anode and the electrochemical process was run with different experimental parameters including counter electrode, electrode gap, applied potential, electrolyte pH, and temperature. Standard samples were determined in the optimum conditions, a linear range of 19.2–11,600 mg l−1 COD and a low detection limit of 0.192 mg l−1 COD were well established with the present approach. The COD value of the simulated organic wastewater determined by this method agreed well with the standard dichromate method, and it showed good accuracy, stability, and reproducibility.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.