Abstract

This article presents a process to calibrate a condensation particle counter (CPC) over the concentration range from 1 particle·cm−3 to 104 particles·cm−3 with traceability to SI units as realized by the National Institute of Standards and Technology (NIST). The process combines two independent steps: a proportionality assessment over the range of the CPC using a consistent diluter and an absolute calibration of the CPC with an aerosol electrometer (AE) at high concentrations. The proportionality assessment, which is the focus of the article, is analogous to the attenuation method for testing the proportionality of high-power laser detectors. This procedure tests the proportionality of one CPC by itself and does not require the use of a calibrated reference CPC. The primary calibration of the CPC with the AE at high concentrations allows for an absolute calibration with NIST traceability, resulting in a simple one-parameter correction to the measured CPC data. The proportionality test enables uncertaintie...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.