The purpose of this study was to determine the concurrent validity of the BOD POD (BP) (Life Measurement Instruments) and Dual Energy X-Ray Absorptiometry (DXA) Elite 4500A (Hologic, Inc.) techniques for assessing the body fat percentage of young women. The participants were forty-three white college-aged women (19.4±1.4 years) with a BMI of 23.4±2.3. Both body composition analyses were completed on the same day and were taken within 10 minutes of each other. Body fat percentage was estimated to be 24.3 (SE=1.1) and 23.8 (SE=0.8) using the BP and DXA techniques, respectively. Exact matches, in terms of body fat percentage, were obtained for 10 of the 43 participants (23.3%). In conclusion, our data supports the concurrent validity of the BP and DXA techniques for assessing body fat in young women. J Am Diet Assoc. 2002;102:1677-1679.