Abstract

Purpose: The purpose of this study was to examine the concurrent validity of a fixated hand-held dynamometer (HHD) in comparison to a load cell in measurement of maximal isometric knee extension force in adults with congenital heart disease (CHD) and healthy adults.Methods: Fourteen adults with CHD and fourteen healthy adults were included. Each participant was tested three times with each method and the mean of the top two results for each participant and method was used in analysis.Results: The agreement between the two methods was excellent in both groups (intraclass correlation coefficient [ICC], 95% confidence interval [CI]) 0.98 (0.92–1.00) in the CHD group and ICC 0.99 (0.96–1.00) in the healthy group). There was a small difference of 19.5 Newton or 4.8% (p<.05) between the two methods in the CHD group. No significant difference was seen between the two methods in the healthy group (p>.05).Conclusions: The fixated HHD demonstrated excellent concurrent validity when compared to a load cell among adults with CHD as well as in healthy adults. Thus, in a healthy population the methods can be used interchangeably, however, a small difference between the methods is seen in the CHD group.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.