Abstract

Proton therapy systems are used worldwide for patient treatment and fundamental research. The generation of secondary particles when the beam interacts with the beamline elements is a well known issue. In particular, the energy degrader is the dominant source of secondary radiation. This poses new challenges for the concrete shielding of compact systems and beamline elements activation computation. We use a novel methodology to seamlessly simulate all the processes relevant to the activation evaluation. A realistic model of the system is developed using Beam Delivery Simulation (BDSIM), a Geant4-based particle tracking code that allows a single model to simulate primary and secondary particle tracking and all particle-matter interactions. The secondary particle fluxes extracted from the simulations are provided as input to FISPACT-II to compute the activation by solving the rate equations. This approach is applied to the Ion Beam Applications (IBA) Proteus®ONE (P1) system and the shielding of the proton therapy research centre of Charleroi, Belgium. Proton loss distributions are used to model the production of secondary neutrals inside the accelerator structure. Two models for the distribution of proton losses are compared for the computation of the clearance index at specific locations of the design. Results show that the variation in the accelerator loss models can be characterised as a systematic error.