Abstract
Customized and standard automated insulin delivery (AID) systems for use in pregnancies of women with preexisting type 1 diabetes (T1D) are being developed and tested to achieve pregnancy appropriate continuous glucose monitoring (CGM) targets. Guidance on the use of CGM for treatment decisions during pregnancy in the United States is limited. Ten pregnant women with preexisting T1D participated in a trial evaluating at-home use of a pregnancy-specific AID system. Seven-point self-monitoring of blood glucose (SMBG) was compared to the closest sensor glucose (Dexcom G6 CGM) value biweekly to assess safety and reliability based on the 20%/20 mg/dL criteria. All participants completed the study with 7 participants satisfying the safety and reliability criteria with a mean absolute relative difference of 10.3%. Three participants did not fulfill the criteria, mainly because the frequency of SMBG did not meet the requirements. Dexcom G6 CGM is safe and accurate in the real-world setting for use in pregnant women with preexisting T1D with reduced SMBG testing as part of a pregnancy-specific AID system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.