Abstract

A new leading-off electrode system, a concentric bipolar electrode, with higher local resolution than other bipolar electrode systems has been developed. To achieve local resolution improvement, criteria for electrode design were demanded. A theoretical study, near-pass filtering, on the decline of the signal amplitude for increasing observation distances in relation to the size of leading-off surfaces of various unipolar and bipolar electrode systems was performed, and recommendations made for real systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.