.Purpose: We explain our concept for customization of a guidance instrument, present a prototype, and describe a set of experiments to evaluate its positioning and drilling accuracy.Methods: Our concept is characterized by the use of bone cement, which enables fixation of a specific configuration for each individual surgical template. This well-established medical product was selected to ensure future intraoperative fabrication of the template under sterile conditions. For customization, a manually operated alignment device is proposed that temporary defines the planned trajectory until the bone cement is hardened. Experiments () with half-skull phantoms were performed. Analysis of accuracy comprises targeting validations and experiments including drilling in bone substitutes.Results: The resulting mean positioning error was found to be at the level of the target point whereas drilling was possible with a mean accuracy of .Conclusion: We proposed a cost-effective, easy-to-use approach for accurate instrument guidance that enables template fabrication under sterile conditions. The utilization of bone cement was proven to fulfill the demands of an easy, quick, and prospectively intraoperatively doable customization. We could demonstrate sufficient accuracy for many surgical applications, e.g., in neurosurgery. The system in this early development stage already outperforms conventional stereotactic frames and image-guided surgery systems in terms of targeting accuracy.

Highlights

  • Accurate positioning and orientation of surgical tools, such as stimulation probes, drills, or biopsy needles, require assistance systems that support the surgeon in performing these demanding tasks

  • On the upper surface of the Trifix, the rigid fixation of the surgical template is ensured by a definite mechanical coupling interface (MCI)

  • We have explored frameless approaches incorporating image-guided surgery” (IGS) systems,[6,7,23] we have opted for the framebased approach in the past due to the prospect of greater accuracy based on experiences with both concepts.[6,24]

Read more

Summary

Introduction

Accurate positioning and orientation of surgical tools, such as stimulation probes, drills, or biopsy needles, require assistance systems that support the surgeon in performing these demanding tasks. This need of assistance applies in particular if the anatomical conditions necessitate an accuracy of a few tenths of a millimeter and if the target region is located invisibly deep inside the body. Rau et al.: Concept description and accuracy evaluation of a moldable surgical targeting system performed in a minimally invasive manner, i.e., replacing the conventional approach by a single drill hole to get access to the inner ear (cochlea). Highly accurate positioning of the drilling tool along a previously planned trajectory is obligatory based on patient-specific imaging

Methods
Results
Discussion
Conclusion