Abstract

A better appreciation of the specific spatial recording characteristics of the single fiber and concentric needle electrode can result in more accurate physiologic and theoretical interpretations of single fiber and quantitative motor unit action potential analysis. We demonstrate by physical modeling that the 90% and 99% amplitude sensitivity envelopes are not simple hemispherical shapes. The 90% sensitivity concentric electrode volume does not extend beyond the insulated portion of the 15 degree beveled surface between the core and cannula and extends only 280 microm perpendicularly from the center of the core's surface. The 99% envelope extends approximately 830 microm perpendicularly from the core's center. This is a much smaller volume of sensitivity than exists for a similarly modeled monopolar electrode. The 90% and 99% envelopes extend to 110 and 320 microm perpendicularly from the exposed single fiber core. Both the single fiber and concentric needle volumes of sensitivity have specific asymmetries described.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.