Abstract

Fluorescence detection is one of the most widely used techniques for measuring analyte concentration in biological applications. Although used frequently with bulk systems, there are important considerations in using this approach in thin (<100 μm) microfluidic lab-on-a-chip (LOC) systems in relating the measured fluorescence signal to analyte concentration. In this work, a general relationship between the photoresponse of an on-chip organic photodetector (OPD) and concentration of a fluorescent dye Rhodamine 6G in a microfluidic chip is presented. The developed theoretical model for photoresponse matches well to the extreme sub-linearity observed in measurements in a dilution series from 10 nM to 1 mM. The influence of the system factors, such as noise, detector sensitivity, and optical power were also analyzed to indicate design changes to improve the system performance and limits of detection even further.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.