Abstract

The aim of this publication is to review the requirements necessary for using computed tomography (CT) for radiotherapy treatment planning. The equipments were tested with different CT phantoms. The authors made several measurements for checking the CT number, the quality and the mechanical parameters of CT tables. The CT numbers measured on the equipments of different manufacturers were in quite good agreement with the IAEA requirements. The geometric distortions of CT images are negligible, while the mechanical parameters of CT tables show considerable variety. A quality assurance - quality control protocol is recommended to implement in a safe workflow.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.