Abstract
Conventional computerized tomography (CT) technique is based on the absorption contrast. In recent years, X-ray phase-contrast CT (PCCT) has been developing rapidly. It uses the phase information to reconstruct the object and provide high contrast and spatial resolution. Diffraction enhanced imaging (DEI) method is one of the three phase-sensitive X-ray imaging techniques. DEI method employs an analyzer crystal to extract the object's refraction information which can be used for CT. However, when DEI refraction CT is combined with the conventional CT algorithm, it should be satisfied that the refraction information of an arbitrary point in the object is invariable at every projection view. In this paper, the invariance condition of refraction information is analyzed and two feasible methods are provided for reconstruction. Using these two methods, two samples of weak absorption are reconstructed with the experimental data obtained at Beijing Synchrotron Radiation Facility (BSRF).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.