Abstract

In the present study a computerized apparatus was employed for scanning light transmission variations along test tubes containing density-separated platelets. The device consists of a stepping motor, a stationary halogen lamp and a photopotentiometer connected to a personal computer. Anticoagulated whole blood was layered on a performed continuous Percoll gradient having a density span from 1090 kg/l (bottom) to 1040 kg/l (top). After centrifugation at 3400g for 1.5 hours, high-density cells (i.e. erythrocytes) pass through to the bottom of the test tube and the lighter platelets remain in the gradient. The test tube is moved by the computer between the halogen lamp and the photopotentiometer. Transmission variations along the gradient were recorded and registered in the computer. Density markers beads were used as an internal standard and platelet peak density was determined. After perforating the test tube the gradient was divided into 45 aliquots. In all fractions determination of platelet counts and mean platelet volume was carried out. In addition, in the aliquots having a platelet count > 20 x 10(12)/l the ratio beta-thromboglobulin per platelet was also determined. The platelet distribution in the gradient was illustrated graphically. A good agreement was found when comparing platelet distributions in the gradients and light transmission variations along the test tubes.