Abstract

An automated flow injection analysis (FIA) system has been developed for the rapid acquisition of liquid/liquid, metal ion distribution ratios (D). The system features automatic switching between aqueous metal sample and wash solutions, on-line solvent extraction, phase separation, and the simultaneous detection of the separated phases by diode-array spectrophotometry. A comparative study of manual, single-stage liquid/liquid extractions with the flow injection system was completed by using a new extraction system UO/sub 2//sup 2 +//benzene/TOPO (trioctylphosphine oxide)/HBMPPT (4-benzoyl-2,4-dihydro-5-methyl-2-phenyl-3H-pyrazol-3-thione). The batch and FIA methods yielded results generally within 5% of each other. The major differences between the two systems are that the FIA system is at least twice as fast, is less labor intensive, is more reproducible, and yields better statistics (a result of the FIA's speed and automation features). Slope analysis of the plotted data from the uranyl extraction studies indicates that the extraction complex is UO/sub 2/(BMPPT)/sub 2/(TOPO).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.