Abstract

To compare the effectiveness of computer-assisted surgery (CAS) and mechanical jig technique in hip resurfacing arthroplasty, we reviewed 176 hip resurfacing arthroplasty performed in 158 patients. The initial 131 hips were resurfaced using the conventional mechanical jig technique, and the remaining 45 hips with the CAS technique. The demographic data of the patients were similar for both techniques. Follow-up radiographs taken 2 months after the surgery showed patients in the CAS technique having a significantly better alignment of the femoral component on the frontal and sagittal planes. There was no difference in the risk of implant notching on the femoral neck for both techniques. The implant sizes were similar for both techniques; thus, the amount of bone stock preserved was not technique specific.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.