Abstract
The purpose of the study was to evaluate the use of the Vogele-Bale-Hohner (VBH) mouthpiece, which is attached to the patient's upper jaw by negative pressure, for patient-image registration and for tracking the patient's head during image-guided neurosurgery. A dynamic reference frame (DRF) is reproducibly mounted on the mouthpiece. Reference points, optimally distributed and attached to the mouthpiece, are used for registration in the patient's absence on the day before surgery. In the operating room, the mouthpiece and DRF are precisely repositioned using a vacuum, and the patient's anatomical structures are automatically registered to corresponding ones on the image. Experimental studies and clinical experiences in 10 patients confirmed repeated (rigid body) localization accuracy in the range of 0 to 2 mm, throughout the entire surgery despite movements by the patient. Because of its noninvasive, rigid, reliable, and reproducible connection to the patient's head, the VBH vacuum-affixed mouthpiece grants the registration device an accuracy comparable to invasive fiducial markers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.