Abstract

The present study describes an automatic procedure for calculating the ejection fraction on the basis of ultrasound images. This procedure is divided into three steps, and programmed in C/UNIX. In an initial step, an algorithm, known as sequence analysis, is used to identify the enddiastolic and endsystolic images from a series of ultrasound images. With the use of a template matching procedure for calculating displacement vectors, a method has been developed for the first time for analysing moving images for identification purposes. The subsequent image analysis employs various filter processes for noise suppression, contrast enhancement, image sharpening, edge detection and segmentation. On the basis of the geometrical data of the segmented left ventricle obtained in this way the respective volumes and ejection fraction are calculated. To validate this automatic procedure, the series of ultrasound images were examined by a cardiologist. It was found that the identification of the endsystolic and enddiastolic image was correct, and that there was good agreement between the automatically determined and manually determined ejection fraction.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.