Abstract

To prospectively evaluate the accuracy and reliability of "freehand" posttraumatic orbital wall reconstruction with AO (Arbeitsgemeinschaft Osteosynthese) titanium mesh plates by using computer-aided volumetric measurement of the bony orbits. Bony orbital volume was measured in 12 patients from coronal CT scan slices using OsiriX Medical Image software. After defining the volumetric limits of the orbit, the segmentation of the bony orbital region of interest of each single slice was performed. At the end of the segmentation process, all regions of interest were grouped and the volume was computed. The same procedure was performed on both orbits, and thereafter the volume of the contralateral uninjured orbit was used as a control for comparison. In all patients, the volume data of the reconstructed orbit fitted that of the contralateral uninjured orbit with accuracy to within 1.85 cm3 (7%). This preliminary study has demonstrated that posttraumatic orbital wall reconstruction using "freehand" bending and placement of AO titanium mesh plates results in a high success rate in re-establishing preoperative bony volume, which closely approximates that of the contralateral uninjured orbit.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.