Abstract This report describes a method for the preoperative determination of radioactive interstitial source placement within computed tomography (CT)-defined tumor boundaries. The method utilizes CT data obtained under stereotactic conditions. Tumor boundaries are digitized from CT slices and are retained in a three-dimensional computer matrix. A solid tumor volume is created by an interpolation program and may be sliced orthogonal to any specific stereotactic surgical view line. The surgeon may simulate radioactive source placement within the slices and view the resultant isodose configuration against tumor contours on successive slices. Once the best source placement has been determined, the computer outputs the mechanical adjustments that will be necessary on a stereotactic frame located in the operating room for the stereotactic placement of each source and gives the length of each source. Sources are stereotactically implanted utilizing a double-catheter afterloading technique.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.