Abstract

Calibration charts for cobalt 60, giving treatment times for field sizes from 4 × 4 cem to 32 × 32 cm have been produced on a computer. After the basic data required are collected by a large number of measurements for different field sizes and depths, the calibration charts can be generated from a single calibration point and are available for 5-mm depth dose, air exposure in roentgens, and tumor dose in increments of 1 cm. Use of these charts makes calculation of equivalent square fields and depth dose interpolation unnecessary. Methods for application to any cobalt therapy unit are discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.