Abstract

This in vitro biomechanical study compared a conventional balancing technique in knee arthroplasty to a technique using computer assistance. The experimental technique used a soft tissue tensioner instrumented with computer-monitored load cells to quantify soft tissue tension. To assess outcome, a tibial load transducer measured medial and lateral compartment forces and tibial rotation. An electromagnetic tracking system quantified knee position under simulated muscle loading. The computer-assisted technique improved knee balance before insertion of components. However, once components were implanted, there was no difference in knee load balance between the 2 techniques. No correlation was shown between compartmental load balance and tibial rotation or mechanical axis misalignment. Although computer-assisted surgery may improve technical accuracy, further work is necessary to achieve an optimal final knee load balance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.