Abstract

The effect of fluoroscopy-based navigation for femoral fracture reduction on the prevention of malrotation was examined in an experimental setting followed by a first case series. Eleven cadaver femurs were used. All femurs were reduced by closed methods. An optoelectronic navigation system was utilized to check for fragment reduction and alignment. Fluoroscopic control without navigation was used as the control group. The Six Sigma Analysis [offset capability index (C (pk)) = 1.3] was used to compare the probability of outliers of more than 15 degrees . In the clinical case series the same navigation tool was used in ten non-consecutive patients with femoral fractures. Torsional differences between both legs were measured postoperatively by CT scan. The highest malrotation in the navigated group was 7.0 degrees for the cadaver testings, while two femurs in the control group showed a difference of more than 10 degrees (10.3 degrees , 17.4 degrees). Only the navigated group showed a sufficient offset capability index (C (pk-navigated) = 1.83; C (pk-conventional) = 0.59). In the clinical series nine femurs were successfully reduced by navigation control. The average malrotation was 6.6 degrees . No patient had a torsional difference of more than 10 degrees. Navigated femoral nailing reduces the risk for outliers of postoperative torsional differences and might avoid revision surgery for malrotation. IV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.