Abstract

The purpose of this study was to investigate a new software program for semiautomatic measurement of the volume and mass of ground-glass nodules (GGNs) in a chest phantom and to investigate the influence of CT scanner, reconstruction filter, tube voltage, and tube current. We used an anthropomorphic chest phantom with eight artificial GGNs with two different CT attenuations and four different volumes. CT scans were obtained with four models of CT scanner at 120 kVp and 25 mAs with a soft and a sharp reconstruction filter. On the 256-MDCT scanner, the tube current-exposure time product and tube voltage settings were varied. GGNs were measured with software that automatically segmented the nodules. Absolute percentage error (APE) was calculated for volume, mass, and density. Wilcoxon signed rank, Mann-Whitney U, and Kruskal-Wallis tests were used for analysis. Volume and mass did not differ significantly from the true values. When measurements were expressed as APE, the error range was 2-36% for volume and 5-46% for mass, which was significantly different from no error. We did not find significant differences in APE between CT scanners with filters for lower tube current for volume or lower tube voltage for mass. Computer-aided segmentation and mass and volume measurements of GGNs with the prototype software had promising results in this study.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.