Abstract

Closed reduction is of great benefit for fracture healing. However, achieving this without sacrificing the reduction accuracy and exposing the surgeon and patient to excessive radiation is difficult. A novel parachute guiding system (ParaEx System) was developed for closed reduction of fractures based on computed tomography data. The system included two counter guides with stainless tubular markers that could be attached to the unilateral external fixator. Comminuted tibial diaphyseal fracture models were used to validate the ParaEx System. The mean errors (and standard deviations) of residual rotational and translational deformity were 0.67° ± 0.45°, 0.92° ± 1.00°, and 0.64° ± 0.50° in rotation and 1.30 ± 1.10 mm, 1.13 ± 0.70 mm, and 0.94 ± 0.92 mm in translation about the X, Y, and Z axes of the local coordinate axes, respectively. The ParaEx System was useful for accurate closed reduction of fractures at low cost.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.