Abstract
A multiple-energy computed tomography (MECT) system that employs monochromatic and tunable 33–100 keV X rays from a superconducting wiggler at the National Synchrotron Light Source is being developed at Brookhaven National Laboratory. The CT configuration is that of a fixed, horizontal fan-shape beam and a subject seated in a rotating chair. Two quantitative CT methods will be used: a) K-edge subtraction of intravenously administered iodine (or a heavier element) to image brain tumors, large blood vessels of the lower head and neck, and arteriovenous malformations; and b) dual photon absorptiometry to obtain two brain CT images that map the low− Z elements and the intermediate− Z elements (i.e. P, S, Cl, K, Ca, and Fe) separately. The system is expected to provide 0.5 mm spatial resolution, horizontally, with unprecedented image contrast and accuracy of quantification. The system will employ a two-crystal monochromator and a high-purity Ge linear array detector.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.