Abstract

In this study, we aimed to compare the split-bolus and single-bolus computerised tomography (CT) urography and determine if this offers a reduction in radiation dose without compromising image quality. A retrospective evaluation was performed on 88 patients undergoing split-bolus CT urography and this was compared to a control group of 101 consecutive patients undergoing single-bolus CT urography. A radiation dose analysis was performed on each subject. Subjects with urinary bladder lesions, hydronephrosis, renal masses or cysts >3 cm in diameter were excluded. All images were classified according to image quality by 2 consultant radiologists. Opacification of the renal parenchyma, pelvicalyceal system, proximal ureters and urinary bladder were comparable between the 2 techniques, whilst image quality of the middle and distal third of the ureters was better using the split-bolus technique. The mean dose length product (DLP) for the single-bolus technique was 1324.1 mGy-cm, whilst that of the split-bolus technique was 885.7 mGy-cm. The mean effective dose reduction was calculated to be 31.1% between the 2 groups. The split-bolus technique gives a reduced radiation dose without compromising image quality. The associated reduction in images is beneficial for data storage and reporting efficiency. As such, our department will adopt the split-bolus technique for young, low-risk patients.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.