Abstract

In the contribution, we present a laboratory system capable of X-ray computed tomography (XCT) scanning of an periodically moving or oscillating object. The system is an in-house developed XCT setup with electromagnetic voice coil actuator mounted on top of the rotary stage of the setup. The strict synchronization of the components, the rotary stage, the electromagnetic actuator movement and the detector readout is accomplished with use of the detector hardware trigger and hard real-time Linux operating system. Cylindrical sample manufactured from epoxy resin with metal particles to enable movement tracking is scanned in a stationary position and during periodical movement induced by the vibration stage. The volumetric data of the scans is compared and the results of this contribution represent an important step towards identification of defects through modal analysis of in-situ harmonically vibrating object.