Abstract

Computed tomography (CT) allows calculation of anatomic and prosthetic humeral head retroversion. The purpose of this study was to demonstrate how the retroversion angle measured by CT scan varied with changes in arm position in the CT scan reference system. A trigonometric analysis shows that the measured retroversion angle decreases when the arm is in extension and increases when it is in flexion, compared with the true retroversion angle determined perioperatively. For the same degree of flexion, or extension, the error of measurement is greater when the initial true retroversion is low. A shoulder prosthesis with 20° of true head retroversion was radiologically scanned with different degrees of flexion, extension, and abduction. The results validate the theoretical analysis. Trigonometric formulas are proposed to correct the retroversion angle measured by CT scan. (J Shoulder Elbow Surg 2001;10:546-9.)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.